Features & Specifications

Cabin Specifications
Flow Volume Measurement: Heated Lilly-style Pneumotachograph
  • Range: 0 to +/- 20 L/sec
  • Accuracy: Better than 2% of scale
  • Resistance: @ 10 L/sec = 12 mmH2O
  • Pressure Transducer: Solid State
  • Frequency Response: Flat to 50 kHz
  • Range: 0 to +/- 100 cmH2O
Environmental Sensors
  • Barometric Pressure - Internal barometer transducer
  • Temperature - Internal spirometer thermistor
Respiratory Pressure MIP/MEP: Solid State Pressure Transducer
  • Sensitivity: 150 cmH2O
  • Resolution: 0.01 cmH2O
  • Linearity: < 0.1%
  • Accuracy: < 0.01%
Cabin Pressure PBox: Solid State Pressure Transducer
  • Sensitivity: 0.5 cmH2O
  • Resolution: 0.0001 cmH2O
  • Linearity: < 0.1%
  • Accuracy: < 0.01%
Mouth Pressure PAlv: Solid State Pressure Transducer
  • Sensitivity: 50 cmH2O
  • Resolution: 0.001 cmH2O
  • Linearity: < 0.1%
  • Accuracy: < 0.01%
Helium Analyzer: Thermal Conductivity
  • Range: 0 to 15.00% He
  • Accuracy: +/- 1% (2 to 14% He)
  • Response Time: Fixed to 30 seconds
  • Oxygen Compensation: Automatic
Carbon Monoxide Analyzer: Chemical Fuel Cell
  • Range: 0 to 0.500% CO
  • Accuracy: +/- 2% (0 to 0.330% CO)
  • Response: Fixed to 30 seconds
  • Working Life: 18 months to 2 years
  • Replacement: Simple plug-in part
Oxygen Analyzer: Galvanic Cell
  • Range: 0 to 100% O2
  • Accuracy: +/- 1% (0 to 100% O2)
  • Response Time: Fixed to 30 seconds
  • Working Life: 18 months to 2 years
  • Replacement: Simple plug-in part
Patient Manifold
  • Pneumotach Cleaning: Easily disassembles for cold sterilization
  • Patient Valve: Pneumatic operation at 50psi
  • Deadspace: < 100ml
  • Valve Support Arm: Fully adjustable
Computer Communication
  • Serial with USB adaptor
Computer Requirements: Minimal Specifications
  • CPU: Intel Core i5 (or better)
  • RAM: 8 GB or greater
  • STORAGE: 1GB for installation, plus 1GB per year for data growth
  • Operating System: Windows® 7 Professional (32/64-bit) or better
  • USB: 2 available USB ports (2.0 or better)
Mobile Cart: 82 x 66 x 60 cm
  • Adjustable Monitor height
  • Adjustable RH or LH keyboard/mouse tray
  • Accessories drawer
  • Power supply and power distribution
  • Power: Single plug 110v
Gas Requirements: Single Cylinder
  • Test Gas: DLCO mixture (0.300% CO, 10.0% He, 21% O2 balance N2)
  • Regulator ‘H’ or ‘E’ provided with system
Test capability

Static Spirometry, Flow Volume Loops, MVV, DLCO, Lung Volumes by N2 recovery, VTG. RAW, MIP/MEP, Cough Peak Flow & Bronchial Challenge.

Optional with Nonin WristOx (Bluetooth Six Minute Walk)

Predicteds

GLI, ATS/ERS, NHANES III, Polgar, Knudson, Crapo & Morris, Hsu, Wang & Dockery anad many more.

Reporting Library
  • Logo and header design
  • Manual entry information
  • Numerous test graphics
  • Summary of all efforts
  • Overlay of graphics
  • Past test result history
  • Serial data trend graphing
  • Computer impressions
  • Captured digital signature
  • Administrative Reports
  • QA Reports
  • Research Query Reports
Interfacing

Health Level Seven (HL7)

Remote Interpretation:

When tests are completed, results can be routed to any network/internet -connected station for physician interpretation

Remote Support

Using secure internet connections