Features & Specifications

Flow Volume Measurement: Rolling Seal Spirometer
  • Volume: 12 L
  • Range: 0 to +/- 15 L/sec
  • Accuracy: Better than 2% of scale
  • Resistance: @ 10 L/sec < 5 mmH2O
  • Digital resolution: Optical encoder 8ml
  • Temperature: Internal spirometer thermistor
DLCO Volume Measurement: Lilly-style Pneumotachograph
  • Flow Range: 0 to +/- 20 L/sec
  • Accuracy: Better than 2% of scale
  • Resistance: @ 10 L/sec < 12 mmH2O
  • Pressure: Transducer Solid State
  • Frequency: Response Flat to 50KhZ
  • Pressure Range: 0 to +/- 100 cmH2O
Respiratory Pressures: Solid State Pressure Transducer
  • Range: 0 to +/- 300 cmH2O
Helium Analyzer: Thermal Conductivity
  • Range: 0 to 15.00% He
  • Accuracy: +/- 1% (2 to 14% He)
  • Response: Time Fixed to 30 seconds
  • Oxygen: Compensation Automatic
Carbon Monoxide Analyzer: Chemical Fuel Cell
  • Range: 0 to 0.500% CO
  • Accuracy: +/- 2% (0 to 0.300% CO)
  • Response: Fixed to 30 seconds
  • Working: Life 2 to 3 years
  • Replacement: Simple plug-in part
Oxygen Analyzer Galvanic Cell
  • Range: 0 to 100% O2
  • Accuracy :+/- 1% (0 to 100% O2)
  • Response Time: Fixed to 30 seconds
  • Working Life: 18 months to 2 years
  • Replacement: Simple plug-in part
Computer Communication
  • Serial with USB adaptor
Computer Requirements: Minimal Specifications
  • CPU: Intel Core i5 (or better)
  • RAM: 8 GB or greater
  • STORAGE: 1GB for installation, plus 1GB per year for data growth
  • Operating System: Windows® 7 Professional (32/64-bit) or better
  • USB: 2 available USB ports (2.0 or better)
Dimensions SpiroAir Unit
  • Spirometer & main chassis: 7" W x 13" H x 20" D
  • Weight: 20 kg (44 lb)
  • Patient Valve Arm: 26" vertical adjustment
Mobile Cart: 82 x 66 x 60 cm
  • Adjustable Monitor height
  • Adjustable RH or LH keyboard/mouse tray
  • Accessories drawer
  • Power supply and power distribution
  • Power: Single plug 110v
Gas Requirements Single Cylinder
  • DLCO Gas: Standard DLCO mixture (0.300% CO, 10.0% He, 21.0% O2 balance N2)
  • Regulators ‘H’ or ‘E’ provided with system
Lung Volume Gases 100% Helium and 100% Oxygen
  • All regulator outputs: 50 psi
Test capability

Static Spirometry, Flow Volume Loops, MVV, DLCO, Lung Volumes by Helium dilution, MIP/MEP, Cough Peak Flow & Bronchial Challenge.

Optional with Nonin WristOx: Bluetooth Six Minute Walk

Predicteds

GLI, ATS/ERS, NHANES III, Polgar, Knudson, Crapo & Morris, Hsu, Wang & Dockery and many more.

Reporting Library
  • Logo and header design
  • Manual entry information
  • Numerous test graphics
  • Summary of all efforts
  • Overlay of graphics
  • Past test result history
  • Serial data trend graphing
  • Computer impressions
  • Captured digital signature
  • Administrative Reports
  • QA Reports
  • Research Query Reports
Interfacing

Health Level Seven (HL7)

Remote Interpretation

When tests are completed, results can be routed to any network/internet -connected station for physician interpretation