Features & Specifications: TransAir3 - Portable

Flow Volume Measurement: Heated Lilly-style Pneumotachograph
  • Range: 0 to +/- 20 L/sec
  • Accuracy: Better than 2% of scale
  • Resistance: @ 10 L/sec = 12 mmH2O
  • Pressure Transducer: Solid State
  • Frequency Response: Flat to 50 kHz
  • Range: 0 to +/- 100 cmH2O
Environmental Sensors
  • Barometric Pressure - Internal barometer transducer
  • Temperature - Internal spirometer thermistor
Respiratory Pressures: Solid State Pressure Transducer
  • Range: 0 to +/- 300 cmH2O
Helium Analyzer: Thermal Conductivity
  • Range: 0 to 15.00% He
  • Accuracy: +/- 1% (2 to 14% He)
  • Response Time: Fixed to 30 seconds
  • Oxygen Compensation: Automatic
Carbon Monoxide Analyzer: Chemical Fuel Cell
  • Range: 0 to 0.500% CO
  • Accuracy: +/- 2% (0 to 0.330% CO)
  • Response: Fixed to 30 seconds
  • Working Life: 18 months to 2 years
  • Replacement: Simple plug-in part
Oxygen Analyzer: Galvanic Cell
  • Range: 0 to 100% O2
  • Accuracy: +/- 1% (0 to 100% O2)
  • Response Time: Fixed to 30 seconds
  • Working Life: 18 months to 2 years
  • Replacement: Simple plug-in part
Computer Communication
  • Serial with USB adaptor
Computer Requirements: Minimal Specifications
  • CPU: Intel Core i5 (equivalent or better)
  • RAM: 2 GB or greater
  • STORAGE: 1GB for installation, plus 1GB per year for data growth
  • Operating System: Windows® XP Professional, 7 or 8 (32/64-bit)
  • USB: One available USB port (2.0 or better)
Dimensions etc.: TransAir3 Unit
  • Weight: 12.2 kg (27 lb)
  • Instrument: 16.4 x 16.3 x 16.7 cm
  • Valve Arm-Span: 26" vertical adjustment
  • Optional Mobile Cart: 82 x 66 x 60 cm
  • Power: Single plug 110v
Platform & Travelling Case

Platform

  • Weight: 4 lb
  • Dimensions: 22 x 18 in

Travelling Case

  • telescoping handle for pulling or pushing on wheels
  • Rubber grip handles
  • push/lift latches
  • customized foam interior
  • Weight: 60 lb (Fully loaded with device, valve with umbilical cord, arm, and platform)
  • Dimensions: 31.3 x 20.4 x 12.2 in
Gas Requirements: Single Cylinder
  • Test Gas: DLCO mixture (0.300% CO, 10.0% He, 21% O2 balance N2)
  • Regulator ‘H’ or ‘E’ provided with system
Test capability

Static Spirometry, Flow Volume Loops, MVV, DLCO, Lung Volumes by N2 recovery, MIP/MEP, Cough Peak Flow & Bronchial Challenge.

Optional with Nonin WristOx (Bluetooth Six Minute Walk)

Predicteds:

GLI, ATS/ERS, Polgar, Knudson, Crapo & Morris, Hsu, Wang & Dockery and many more.

Reporting Library
  • Logo and header design
  • Manual entry information
  • Numerous test graphics
  • Summary of all efforts
  • Overlay of graphics
  • Past test result history
  • Serial data trend graphing
  • Computer impressions
  • Captured digital signature
  • Administrative Reports
  • QA Reports
  • Research Query Reports
Interfacing:

Health Level Seven (HL7)

Remote Interpretation

When tests are completed, results can be routed to any network/internet -connected station for physician interpretation

Remote Support

Using secure internet connections