ComPAS2 Release Build 2023.1.3 is now generally available for customers. This build includes the enhancements from all prior ComPAS2 releases as well as the following updates.

Fixed

  • [#7232] Graph Editor choose variables screen for serial graphs does not display any variables for non-US regions
  • [#7234] CPX import is broken due to composite breath TestVars getting populated with variables from TestSummary
  • [#7239] FVC and SVC grades need to be F if calibrations didn’t pass or were out of date
  • [#7253] Group names in the Configuration/Active Directory screen now display properly
  • [#7258] ComPAS1 to ComPAS2 Converter needs to pull the appropriate values for DLCO depending on the ComPAS2 configuration setting for Traditional or SI units
  • [#7261] Summary data can be overwritten by efforts that have a variable that shouldn’t exist in its collection inadvertently added
  • [#7263] Image on the plethysmography metronome view gets cut off when the view is resized too small
  • [#7316] Fix inability to save a test that has a Clinical Alert message with a length greater than 255
  • [#7328] Technician notes disappearing if notes are entered and you navigate away from the results screen and then navigate back
  • [#7337] COM port does not always get properly cleaned up after failing a flow calibration
  • [#7339] Attempting to save a flow calibration or span when there are no strokes causes the calibration screen to remain open but unusable
  • [#7342] Ducharme AOS z-scores are calculating incorrectly
  • [#7372] FVC efforts that should have a U (usable) grade are being checked for repeatability and achieving higher grades in some cases

Changed

  • [#7245] Change the increments for height and weight up/down controls (cm = 0.1, m = 0.001, kg = 0.1)
  • [#7259] Allow DLCO [Unc] and DLCO [Cor] to have zero or negative values
  • [#7264] BTPS equation for non-heated pneumotach devices should use the environmental temperature for exhaled temperature
  • [#7284] Refresh device calibrations when device changes, or when navigating to the calibrations or results screens
  • [#7351] Improve the alignment of gas signals to flow during Fast Gas DLCO

If you are interested in learning more about ComPAS2 updates or would like to schedule an update for your facility, contact Morgan Scientific Technical Support.