ComPAS2 Release Build 2023.1.5 is now generally available for customers. This build includes the enhancements from all prior ComPAS2 releases as well as the following updates.

Fixed

  • [#7583] MedGraphics CPX importer can misalign imported columns if the column headers have combined VarName and Unit in the form “VarName (Unit)” but a column is missing the unit
  • [#7684] If a test protocol is started after baseline efforts have already been performed, choosing “Skip” will go straight to Level 1 and skip Diluent

Changed

  • [#7629] Add a watchdog for the VitaloLAB / QUB that samples the patient valve for 2 minutes after each effort to flush it with room air
  • [#7631] In Fast Gas DLCO align the start of the gas drop to the start of exhale
  • [#7633] Add a configuration setting for washout volume to Fast Gas DLCO
  • [#7667] Fast Gas DLCO: Fail ATS if the effort has evidence of a leak or Valsalva maneuver
  • [#7674] Change the target exhale flow meter ranges during Fast Gas DLCO
  • [#7212] Automatically restart ComPAS2 Windows Services when changes made to database configuration file
  • [#7587] Update FDA UDI for 2023.1.5

If you are interested in learning more about ComPAS2 updates or would like to schedule an update for your facility, contact Morgan Scientific Technical Support.

Return to the Technical Bulletin Main Page, or review more recent bulletins: