ComPAS2 Release Build 2023.1.5 is now generally available for customers. This build includes the enhancements from all prior ComPAS2 releases as well as the following updates.
Fixed
- [#7583] MedGraphics CPX importer can misalign imported columns if the column headers have combined VarName and Unit in the form “VarName (Unit)” but a column is missing the unit
- [#7684] If a test protocol is started after baseline efforts have already been performed, choosing “Skip” will go straight to Level 1 and skip Diluent
Changed
- [#7629] Add a watchdog for the VitaloLAB / QUB that samples the patient valve for 2 minutes after each effort to flush it with room air
- [#7631] In Fast Gas DLCO align the start of the gas drop to the start of exhale
- [#7633] Add a configuration setting for washout volume to Fast Gas DLCO
- [#7667] Fast Gas DLCO: Fail ATS if the effort has evidence of a leak or Valsalva maneuver
- [#7674] Change the target exhale flow meter ranges during Fast Gas DLCO
- [#7212] Automatically restart ComPAS2 Windows Services when changes made to database configuration file
- [#7587] Update FDA UDI for 2023.1.5
If you are interested in learning more about ComPAS2 updates or would like to schedule an update for your facility, contact Morgan Scientific Technical Support.